1250  Analizador de Respuesta de Frecuencia                                                                

 
The 1250 FRA is an instrument which is capable of meeting today's measurement requirements for dynamic system analysis. Using the single sine measurement technique it provides precise data relating to the gain and phase between any points in a dynamic system. This instrument enables researchers to perform searching analysis to assess performance or characterize both simple or complex systems.
 
The device under test is stimulated by a sinewave generated by the 1250 FRA and the response analyzed at one or more points in the system. These responses are then correlated with the applied stimulus to determine the amplitude and phase relative to the generator. The ratio of the two signals can then be calculated to provide the system transfer function.
 
The process rejects all harmonics and, by increasing the integration time even signals buried in noise can be accurately measured.
The 1250 FRA has been used to engineer and optimize a variety of performance parameters for measurement systems including:
• Servo systems
• Hydraulic actuators
• Electromechanical components
The system has the following options available:
• Sincronizador - Opción 12501
• Modulador Demodulador - Opción 12502
Sincronizador
If it is required to measure the phase shift in the carrier system itself, then it is necessary that the generator of the frequency response analyser is locked to the carrier in both frequency and phase. This is accomplished using the optional synchroniser, which operates over the full frequency range of the instrument. The synchroniser also enables the generator to be locked to a rotating or reciprocating component, and measurements can then be made relative to this.
Modulador Demodulador
The optional modulator/demodulator interfaces the 1250 to ac carrier systems. Individual switching of either the modulator or demodulator, and full control of the analysis channels that are being demodulated, ensure the multi-channel flexibility. The response of the systems ac components, dc components and indeed the integral system modulators and demodulators can all be measured at the same time.

 

Generador

 

 

 

 

Frequencia

 

 

 

 

- Rango

10µHz to 65kHz

 

 

 

- Resolución

1 in 65535

 

 

 

- Error

<0.01%

 

 

 

- Sweep

Logarithmic, up or down linear, up or down harmonic

 

 

 

Amplitud

 

 

 

 

- Range

10mV to 10.23V (triangle: 5.11V)

 

 

 

- Resolution

1 in 1023

 

 

 

- Error

<1% ± 1 digit

 

 

 

Bias - Polarización

 

 

 

 

- Rango

-10.23 to +10.23V

 

 

 

- Resolution

1 in 1023

 

 

 

- Error

<1% ± 1 digit

 

 

 

Distortion

<2%

 

 

 

Maximum output

 

 

 

 

- Bias + signal

15V peak

 

 

 

- 1250N

20V peak

 

 

 

Output impedance

50W+2%

 

 

 

- 1250N

<1W

 

 

 

Maximum voltage, Lo to ground

150V

 

 

 

Impedance, Lo to ground

100kW, 100pF

 

 

 

Stop control

Immediate or at 0º, 90º, 180º, 270º

 

 

 

Stop input

Contract closure or TTL logic 0

 

 

 

Connection

 

 

 

 

- Front

Floating, 4mm

 

 

 

- Rear

Floating, BNC

 

 

 

Output is short-circuit proof

 

 

 

 

Analizadores (el 1250 dispone de 2 y el 1254 dispone de 4 Analizadores)

Rango Nominal

Sensibilidad

Full scale pk input

Com. mode rejected

30mV

1uV

45mV

30V

300mV

10uV

500mV

30V

3V

100uV

5V

30V

30V

1mV

50V

500V

300V

10mV

500V

500V

Sensitivity is for integration time >100ms

 

 

Maximum input, Hi or Lo to ground

500V peak, 300V rms

Coupling

dc or ac (<1dB at 2.5Hz)

Configuración de entrada

 

Connection

 

- Front

differential, 4mm

- Rear

differential, BNC

Impedance, Hi or Lo to ground

1MW

- Front sockets

<70pF

- Rear sockets

<100pF

Common mode rejection, dc coupling, up to 100Hz

 

- Up to 50V peak

>65dB

- Over 50V peak

>60dB

Cross channel isolation, 1kW across inputs

 

- Up to 10kHz

>100dB

Integration time

 

- Minimum

the longer of 1 cycle or 10ms

- Maximum

10^6 cycles or 10^5s

Auto-integration

 

- Minimum

the longer of 3 cycles or 1.5s

- Maximum

the programmed integration time

Variance in results, at 90% confidence

 

- Long integration (signal >0.02% range)

<1%

- Short integration (signal >0.2% range)

<10%

Measurement delay

Variable from zero to 10^6 cycles or 10^5s

 

 

 

 

Sincronizador - Opción 12501

 

 

 

Input configuration

 

 

 

 

Connection

differential, rear terminals

 

 

 

Coupling

dc or ac (<3dB at 3Hz)

 

 

 

Impedance, Hi or Lo to ground

>200kW, <100pF

 

 

 

Common mode rejection, dc coupling up to 100Hz

>50dB

 

 

 

- Maximum rejected

20

 

 

 

Maximum input, Hi or Lo to ground

350V peak, 250V rms

 

 

 

Sincronización

 

 

 

 

Frequency range

1mHz to 65kHz

 

 

 

Sensitivity

0.25V

 

 

 

Level adjustment

±5V in steps of 0.02V

 

 

 

Time to synchronise

the longer of 4 cycles or 500ms

 

 

Opción Modulador/Demodulador -12502

 

 

 

Input configuration

 

 

 

Connection

differential, rear terminals

 

 

Coupling

ac

 

 

Impedance, Hi or Lo to ground

>100kW, <100pF

 

 

Common mode rejection, up to 100Hz

>50dB

 

 

Maximum common mode

300V

 

 

Maximum input, Hi or Lo to ground

350V peak, 250V rms

 

 

Portadoras 1 and 2

 

 

 

Frequency range

48Hz to 20kHz

 

 

Voltage range

6V to 250V rms

 

 

Salida Generador

 

 

 

May modulate either carrier 1 or carrier 2

 

 

 

Carrier phase shift

 

 

 

- 50Hz to 300Hz

<3º

 

 

- 300Hz to 3kHz

<1º

 

 

- 3kHz to 20kHz

<6º

 

 

Analizadores

 

 

 

Either carrier may demodulate any analyser

 

 

 

Quadrature rejection

 

 

 

- 1250 y 1254

>26dB

 

 

- 1251

>14dB

 

 

Additional errors when demodulating

Mod frequency = 0.05xcarrier

 

 

Input >10% full scale, integration time

200ms

 

 

- r

<0.5%

 

 

- log r

<0.05dB

 

 

- q, single channel

<0.5º

 

 

- q, point to point

<1º

 

 

 

· Ver manual de Operaciones y Especificaciones

· Solicitar información sobre este producto

 

 

 

 

Equipos  Instrumentación  y  Control  S.L.

Urbanización Prado Norte, c/ Fresno  1 – bis  28110 Algete – MADRID

Telf.: 91 628 0817  Fax: 91 628 0453  Email: info@eiccontrol.com