1290 Amplificador de Potencia

 

Características Principales

Max applied polarization level 50V / 25A

Test techniques include Pstat / Gstat, ramps, cyclic

voltammetry, pulse tests, ohmic drop, impedance

Frequency Range 10uHz to 10kHz

Impedance Range <1mohm to >100kohm

Compatible with Solartron potentiostats including

1280B, 1285, 1287, 1470 and 1480

Compatible with Solartron FRAs including 1250, 1253,

1255, 1260 (but system must include a potentiostat)

 

Software de PC

CellTest - Battery, Fuel Cell test program provides

charge / discharge cycling, pulse, impedance

CorrWare - DC, cyclic voltammetry

ZPlot (with FRA) – Impedance

 

Funciones

Wide polarization range (50V / 25A) for studies of a wide variety of energy storage devices.

Boosts potentiostat voltage and current - potentiostat voltage is multiplied by 10 and current is

multiplied by 10,000 (maximum voltage 50V, maximum current 25A)

4-terminal measurements essential for accurate measurements of low impedance devices such as

batteries and fuel cells.

10uHz - 10kHz frequency range allows impedance characterisation of cells (series resistance,

diffusion, electrode investigations, etc.)

Pulse techniques - for analysis of series resistance components in energy sources.

Remote control from the potentiostat – automatic switch on and off during the experiment is

controlled by the potentiostat. For 1470 and 1480, the 14701A option must be fitted for the booster to

operate correctly. In the case of 1280B, 1285, 1287, these instrument need to have the remote

control board fitted. This board is now fitted as standard but for all older units, a retrofit will be

required. The unit will need to be returned to service for this to be fitted.

The 1290 is not compatible with 1286, 1284, 1280A and 1287s older than mod strike A10 (mid-1995)

 

Aplicaciones y Usuarios

Test types - Pstat / Gstat DC Polarization scans, cyclic voltammetry, impedance

Applications - Battery / Fuel Cell / Ultra-capacitor charge - discharge cycling, series resistance tests

and impedance tests, Electroplating, Etching , Electrochemistry

 

Usuarios Típicos

Universities (electrochemistry research departments), automotive manufacturers (new 42V batteries

for cars), aerospace / military (power sources for satellites, missiles, torpedoes etc), electric vehicle

power source manufacturers, national labs, battery / fuel cell / supercapacitor research labs.

1290 Power Booster - 50V / 25A lots of

power for testing batteries, fuel cells,

ultra-capacitors...

 

Accesorios del Sistema

Compatible Electrochemical / Energy Storage Test Systems

12608W - Full electrochemical test system (1MHz freq. range) 1287, 1260, ZPlot and Corrware

12558WB - Electrochemical system (1MHz freq. range) 1287, 1255B, ZPlot and Corrware

12508W - Intermediate echem system (65kHz freq. range) 1287, 1250B, ZPlot and Corrware

1280Z - Electrochemical system (20kHz freq. range)

147055B - Multi-channel Battery Test System (1MHz range) 1470, 1255, CellTest

148055B - Multistat Test System (1 MHz range) 1480, 1255, CellTest, Multistat ZPlot / CorrWare

Nota 1: For 1470 and 1480, the 14701A option must be fitted (controls the booster)

Nota 2: when the 1290 Power Booster is used the frequency bandwidth is reduced to 10kHz

for the above systems.

 

 

 